Stainless Steel Tube 2x3

Diameter: 5M

Maintenance and replacement:

  • Check for Buildup: During regular maintenance intervals, visually inspect the stainless steel tube for any visible deposits, soot, or corrosion.
  • Surface Inspection: Inspect the outer surface for signs of heat damage or corrosion, which could compromise the tube's performance over time.

Signs of replacement:

  • Corrosion or Contamination: Over time, high-temperature combustion processes can lead to the buildup of soot, carbon deposits, or even corrosion, especially when analyzing samples that release reactive gases.
  • Physical Damage: If the stainless steel tube becomes bent, dented, or damaged during handling or cleaning, it should be replaced to avoid leaks or blockages in the gas pathway.
  • Clogging or Blockages: If there is a noticeable restriction in gas flow or increased resistance in the system, the stainless steel tube may be clogged with residue and needs cleaning or replacement.

Compatibility: